TTi™Colorimeter SeriesQuick-Start GuideTaylor Technologies, Inc. 800-TEST KIT (837-8548)www.taylortechnologies.com
6 TTi™ Colorimeter Series Quick-Start Guide1. Start Test – Select to access FAVORITES, RECENT TESTS, or ALL TESTS menus.2. Access Test Data – Select t
7TTi™ Colorimeter Series Quick-Start GuideBATTERY INSTALLATION OR CONNECTION TO A POWER SUPPLYSelect one of the following options to supply power to t
8 TTi™ Colorimeter Series Quick-Start GuideINITIAL START-UP AND OPERATIONThe following Quick-Start Test example will walk you through initial start-up
9TTi™ Colorimeter Series Quick-Start GuideSTEP 4*Select Start Test from the MAIN MENU using ; then press ENTER .*Step 4 required for initial start-
10 TTi™ Colorimeter Series Quick-Start GuideSTEP 6Confirm test name and range and sample cell size used in the test procedure. Select a chemical form
11TTi™ Colorimeter Series Quick-Start GuideSTEP 8Zero the Colorimeter Select ZERO using ; then press ENTER .Zero will be displayed.STEP 9Prepare Sa
12 TTi™ Colorimeter Series Quick-Start GuideSTEP 10Access TimerSelect TIMER using ; then press ENTER .STE P 11Start TIMER and AUTO OptionSelect STA
13TTi™ Colorimeter Series Quick-Start GuideSTEP 12Read the Sample When the timer beeps, the instrument will automatically read the sample and the re
14 TTi™ Colorimeter Series Quick-Start GuidePC APPTo use your TTi™ Colorimeter to the fullest extent possible, it is necessary to install a special fr
15TTi™ Colorimeter Series Quick-Start Guide• Business name• Business mailing address• Contact’s telephone number• Contact’s e-mail – where you can
16 TTi™ Colorimeter Series Quick-Start GuideWARRANTY INFORMATIONInstruments in the TTi™ Colorimeter Series found to be defective within five years fro
17TTi™ Colorimeter Series Quick-Start GuideMAINTENANCEDo not use chemicals, solvents, or abrasives to clean or dry any part of the instrument. Clean a
18 TTi™ Colorimeter Series Quick-Start GuideIF YOU NEED TECHNICAL ASSISTANCEContact: Taylor Technologies, Inc. 31 Loveton Circle Sparks, Maryland 2115
19TTi™ Colorimeter Series Quick-Start GuideINSTRUMENT SPECIFICATIONSPerformancePhotometric Range 0–2 ABSPhotometric Accuracy ±0.005 ABS @ 1.0 ABS nomi
20 TTi™ Colorimeter Series Quick-Start GuideInstrument Rating (cont’d)Operational Humidity Limit 90% RH @ 122°F/50°C (non-condensing)Dust and Water I
21TTi™ Colorimeter Series Quick-Start GuideINSTRUMENT CERTIFICATION AND INFORMATION TO THE USERThe TTi™ Colorimeter Series is certified to the follow
22 TTi™ Colorimeter Series Quick-Start Guide• This device complies with Part 15 of the FCC Rules. Operation is subject to the following two condition
23TTi™ Colorimeter Series Quick-Start GuideNOTES______________________________________________________________________________________________________
Part #5543
iTTi™ Colorimeter Series Quick-Start GuideTABLE OF CONTENTSPRECAUTIONS ...2COMPONENTS AND
1TTi™ Colorimeter Series Quick-Start GuideThis Quick-Start Guide provides a brief summary of the TTi™ Colorimeter Series User’s Manual. It is designed
2 TTi™ Colorimeter Series Quick-Start GuidePRECAUTIONSFor operator safety and to avoid damage to the instru-ment, the following precautions should be
3TTi™ Colorimeter Series Quick-Start GuideCOMPONENTS AND OPTIONAL ACCESSORIESThis kit contains the following components:Note: All components and opti
4 TTi™ Colorimeter Series Quick-Start GuideGUIDE TO PARTS, OPERATING CONTROLS, AND MAIN MENUReview the following illustrations to become familiar with
5TTi™ Colorimeter Series Quick-Start GuideBottom View 1. Product Label, shows model, part, and serial numbers 2. Timer Buzzer Opening 3. Battery Compa