16712. Error Codes6.4 Select the sterilization programSelect theappropriate sterilization temperatureand cycle. (For additionalinformation, seeappen
1586.7 Starting the sterilization program.Appendix 2Press the STARTbutton, the machine will beginthe sterilization cycle and complete all phases(fill
14910.5 Replacement of the door sealInstructions manualInstructions manual7. Draining the water tanks1.Connect the supplied drain hose tothe drain v
10138 Advanced adjustment settings8.1 Adjustment of sterilization time and drying time8.1.1 Hold the UNWRAPPED button for 30seconds. The machine will
1112Example of the print reportProgram: wrappedTemperature: 134Pressure: 39 PSIVacuum Num.: 3Dry Time: 15minSter Time:
1112Example of the print reportProgram: wrappedTemperature: 134Pressure: 39 PSIVacuum Num.: 3Dry Time: 15minSter Time:
10138 Advanced adjustment settings8.1 Adjustment of sterilization time and drying time8.1.1 Hold the UNWRAPPED button for 30seconds. The machine will
14910.5 Replacement of the door sealInstructions manualInstructions manual7. Draining the water tanks1.Connect the supplied drain hose tothe drain v
1586.7 Starting the sterilization program.Appendix 2Press the STARTbutton, the machine will beginthe sterilization cycle and complete all phases(fill
16712. Error Codes6.4 Select the sterilization programSelect theappropriate sterilization temperatureand cycle. (For additionalinformation, seeappen
24APPENDIX 4Instructions manualInstructions manualV1:Air release valveV4:Water release valveHYDRAULIC DRAWINGChamberPressuresensorSafety valvePumpCond
6Instructions manual6. Operation6.1 Power on6.1.1 Open the chamber door and remove all inner packing andaccessories.6.1.2 Plug the power cord into
55.1 Pressure display window.Displays the current pressure (PSI)inside the chamber5. Control panelStateCycleInstructions manual5.2 Temperature display
1944. InstallationDIAGRAMS OF THE STERILIZATION PROGRAMMESAPPENDIX 2PROGRAMTemperature(°C)Pressure(PSI)Holding time(min.)Total time(min.)TYPEMAXLOAD(k
203. Packing ContentInstructions manualInstructions manualQuantityAccessoryNo234567891011Loading trayTray rackTray handling toolDoor adjustment toolDr
2122. Technical SpecificationsInstructions manualInstructions manual30015Pressure psiWRAPPEDTime (min)15030Pressure psiWRAPPEDTime (min)(1)Chamber Dim
221Instructions manual9754321681 Distilled (clean) water tank2 Display3 Control panel5 F674 Power switchusesDrain connector, clean water tankDoor hand
TABLE OF CONTENTS1. GENERAL ------------------------------------------------------------ 12. TECHNICAL SPECIFICATIONS---------------------------------
24APPENDIX 4Instructions manualInstructions manualV1:Air release valveV4:Water release valveHYDRAULIC DRAWINGChamberPressuresensorSafety valvePumpCond
TABLE OF CONTENTS1. GENERAL ------------------------------------------------------------ 12. TECHNICAL SPECIFICATIONS---------------------------------
221Instructions manual9754321681 Distilled (clean) water tank2 Display3 Control panel5 F674 Power switchusesDrain connector, clean water tankDoor hand
2122. Technical SpecificationsInstructions manualInstructions manual30015Pressure psiWRAPPEDTime (min)15030Pressure psiWRAPPEDTime (min)(1)Chamber Dim
203. Packing ContentInstructions manualInstructions manualQuantityAccessoryNo234567891011Loading trayTray rackTray handling toolDoor adjustment toolDr
1944. InstallationDIAGRAMS OF THE STERILIZATION PROGRAMMESAPPENDIX 2PROGRAMTemperature(°C)Pressure(PSI)Holding time(min.)Total time(min.)TYPEMAXLOAD(k
55.1 Pressure display window.Displays the current pressure (PSI)inside the chamber5. Control panelStateCycleInstructions manual5.2 Temperature display
6Instructions manual6. Operation6.1 Power on6.1.1 Open the chamber door and remove all inner packing andaccessories.6.1.2 Plug the power cord into