Techn ologyMadeEasy... Instruction Manual PC 10 Waterproof Hand-held pH/Conductivity/Temperature Meter 68X230401 Rev 5 08/04
Instruction Manual PC 10 - 6 - 4. CALIBRATION 4.1. Important Information on Meter Calibration When you recalibrate your meter, old pH and conducti
Instruction Manual PC 10 - 7 - 4.2. Preparing the Meter for Calibration The pH/conductivity/temperature probe included with this meter is designed
Instruction Manual PC 10 - 8 - 4.3. Temperature Calibration The built-in temperature sensor included in the probe is factory calibrated. Calibrate
Instruction Manual PC 10 - 9 - 6. Press the ENTER key to confirm temperature calibration. The CON indicator will flash to confirm calibration. See
Instruction Manual PC 10 - 10 - 4.4. pH Calibration This instrument is capable of up to 3-point pH calibration to ensure accuracy across the entire
Instruction Manual PC 10 - 11 - 6. After the reading has stabilized and the READY indicator is displayed, press ENTER to confirm calibration. A co
Instruction Manual PC 10 - 12 - 4.5. Conductivity Calibration The Important Information on Conductivity Calibration Your meter has four measuring ra
Instruction Manual PC 10 - 13 - 4.6. Calibrating for Conductivity: 1. Pour out two separate portions of your calibration standard and one of DI wat
Instruction Manual PC 10 - 14 - 6. Press the ENTER key to confirm calibration. Upon confirmation, the CON indicator appears briefly. The meter auto
Instruction Manual PC 10 - 15 - 5. MEASUREMENT The READY indicator appears on the display when the readings stabilise. It will turn off if the readi
http://www.novatech-usa.com/Products/Laboratory-Equipment-Supply-Products Tel: (281) 359-8538 Toll Free:(866) 433-6682
Instruction Manual PC 10 - 16 - 6. HOLD FUNCTION This feature lets you freeze the value of the pH or conductivity reading for a delayed observation.
Instruction Manual PC 10 - 17 - 7. PROBE CARE AND MAINTENANCE Under typical operating conditions, the probe will need to be replaced every 6 to 12 m
Instruction Manual PC 10 - 18 - 8. TROUBLESHOOTING Problem Cause Solution No display when turned on 1. Batteries not in place. 2. Batteries not i
Instruction Manual PC 10 - 19 - 9. ERROR MESSAGES LCD Display Indicates Cause Solution Err annunciator Unrecognised input from keypad Wrong inpu
Instruction Manual PC 10 - 20 - 10. SPECIFICATIONS Mode pH Temperature Conductivity Range 0.00 to 14.00 pH 0.0 to 100.0 °C 0 to 19.99 μS 0 to 199.
Instruction Manual PC 10 - 21 - 11. ACCESSORIES 11.1. Replacement Meter and Meter accessories Eutech Instruments Item Ordering Code Waterproof PC 10
Instruction Manual PC 10 - 22 - Oakton Instruments Item Ordering Code Waterproof PC 10 pH / Conductivity / Temperature meter and electrode (EC-COMBI
Instruction Manual PC 10 - 23 - 11.2. Calibration Solutions Note: pH buffer solutions have ±0.01 pH accuracy at 25 °C. Conductivity standard solutio
Instruction Manual PC 10 - 24 - 12. WARRANTY This meter is supplied with a three -year warranty, six-month warranty for probe against significant dev
Instruction Manual PC 10 - 25 - 13. RETURN OF ITEMS Authorisation must be obtained from our Customer Service Department or authorised distributor bef
PrefaceThis manual serves to explain the use of the Waterproof PC 10 hand-held meters. It functions as a step by step guide to help you operate the m
TABLE OF CONTENTS 1.INTRODUCTION ...12. DISPLAY AND KEYPAD FUNCTI
Instruction Manual PC 10 - 1 - 1. INTRODUCTION Thank you for selecting the Waterproof PC 10 portable meter. The meter is a microprocessor-based inst
Instruction Manual PC 10 - 2 - 2. DISPLAY AND KEYPAD FUNCTIONS 2.1. Display The LCD has a primary and secondary display. x The primary display sho
Instruction Manual PC 10 - 3 - 2.2. Keypad The large membrane keypad makes the instrument easy to use. Each button, when pressed, has a corresponding
Instruction Manual PC 10 - 4 - 3. PREPARATION 3.1 Inserting the Batteries The meter is packaged with 4 “AAA” alkaline batteries required for operat
Instruction Manual PC 10 - 5 - 3.2 Connecting the Probe The Waterproof PC 10 meter uses a special combination pH / Conductivity / Temperature probe.