© Adam Equipment Company 2007 Adam Equipment CBK SERIES (P.N. 8509, Revision A7, August 2007) Software Rev.: 4.28 & above
© Adam Equipment Company 2007 8 5.0 DISPLAY 6.0 SYMBOLS AND INDICATORS The LCD has unique symbols to indicate the following: 0 The di
© Adam Equipment Company 2007 9 7.0 BATTERY • The scales can be operated from the battery, if desired. The battery life is approxima
© Adam Equipment Company 2007 10 10.0 OPERATIONS 10.1 ZEROING • You can press the [Zero] key at any time to set the zero point fr
© Adam Equipment Company 2007 11 NOTE: When the container is removed a negative value will be shown. If the scale was tared just before removing
© Adam Equipment Company 2007 12 • If a container is to be used, place this container on the platform before entering parts counting a
© Adam Equipment Company 2007 13 • To count a different sample quantity, pre
© Adam Equipment Company 2007 14 3. If Limit Password is enabled then enter the password which will allow you to change the limits or the operat
© Adam Equipment Company 2007 15 10.6 LIMITS STORING AND RECALLING The scale can store up to 10 sets of high and low limits in memory along with
© Adam Equipment Company 2007 16 NOTE: 1. If the recalled limit is for parts counting, the display will show the last sample value used,
© Adam Equipment Company 2007 17 if 3500g is placed on the scale and percent weighing is selected, the display will show 100.00%. Rem
© Adam Equipment Company 2007
© Adam Equipment Company 2007 18 NOTE: 1. The accumulation function is available only during weighing. It is disabled during parts cou
© Adam Equipment Company 2007 19 • To view the total in memory press the [Print] key when the scale is at zero. The display will show the total
© Adam Equipment Company 2007 20 11.0 RS-232 INTERFACE The CBK Series are supplied with bi-directional RS-232 interface as standard
© Adam Equipment Company 2007 21 Data Format-Parts Counting Output: Weight, Unit weight and number of parts are printed. Data Format- Memory Re
© Adam Equipment Company 2007 22 NOTE: 1. The accumulated total will not be sent to the RS-232 when the continuous print is turned on.
© Adam Equipment Company 2007 23 12.0 CALIBRATION • The CBK scales are calibrated using kilogram weights and CBKa scales are calibrated using
© Adam Equipment Company 2007 24 13.0 PARAMETER SETTINGS Pressing the [Func] key allows the user to access the parameters for customi
© Adam Equipment Company 2007 25 Parameter Description Options Default setting F1 PCt This parameter allows the user to enter the Percen
© Adam Equipment Company 2007 26 13.2 RS-232 PARAMETERS • Shortcut to enter this group is to press and hold the [Print] key for 4 secon
© Adam Equipment Company 2007 27 Scale will perform the following, depending on the Accumulation and Print Settings: ACCUMULATION SETTINGS PRINT
© Adam Equipment Company 2007 1 CONTENTS 1.0 INTRODUCTION ...
© Adam Equipment Company 2007 28 S2 bL Backlight set to always on, always off or automatic on whenever a weight is placed or a key is pressed EL
© Adam Equipment Company 2007 29 14.0 ERROR MESSAGES During the initial power-on testing or during operation, the scale may show an error messa
© Adam Equipment Company 2007 30 15.0 REPLACEMENT PARTS AND ACCESSORIES If you need to order any spare parts and accessories, contact your suppli
© Adam Equipment Company 2007 31 17.0 WARRANTY INFORMATION Adam Equipment offers Limited Warranty (Parts and Labour) for the components fa
© Adam Equipment Company 2007 32 18.0 APPENDIX
Manufacturer’s Declaration of Conformity This product has been manufactured in accordance with the harmonised European standards, following th
ADAM EQUIPMENT is an ISO 9001:2000 certified global company with more than 35 years experience in the production and sale of electronic weighing eq
© Adam Equipment Company 2007 2 1.0 INTRODUCTION • The CBK range of scales provides an accurate, fast and versatile general purpose weighing
© Adam Equipment Company 2007 3 2.0 SPECIFICATIONS CBK 4 CBK 8H CBK 8 CBK 16 CBK 32 CBK 48 Kilograms Maximum Capacity
© Adam Equipment Company 2007 4 OTHER SPECIFICATIONS Units of measure CBK Series- kg, g CBKa Series- kg, g, lb, oz, lb:oz Interface RS-232 b
© Adam Equipment Company 2007 5 3.2 LOCATING • The scales should not be placed in a location that will reduce the accuracy. • A
© Adam Equipment Company 2007 6 3.2 SETTING UP • The CBK Series comes with a stainless steel platform packed separately. • Place t
© Adam Equipment Company 2007 7 4.0 KEYPAD KEYS PRIMARY FUNCTION SECONDARY FUNCTION [Zero] Sets the zero point for all subsequent